116
Symbols, display messages and troubleshooting
117
Symbols, display messages and troubleshooting
9
Symbol... Meaning:
You are requested to run a quality
control test.
This result is/was a quality control result.
The temperature during testing was at
the limit of the permitted range (between
+6 °C − +10 °C or between +40 °C
− +44 °C).
Do not use the result of this test as a
basis for therapeutic decisions.
Turn the meter off. Let the meter
warm up or cool down slowly. Turn
the meter on again and repeat the
test.
During testing the permitted tem-
perature range is from +10 °C to
+40 °C.
Symbol... Meaning:
When this result was saved, the tem-
perature was at the limit of the permit-
ted range.
Do not use the result of this test as a
basis for therapeutic decisions.
The batteries are running low. When
the battery symbol first appears in the
display, you can perform about 50 more
tests. Continue testing until all the test
strips in the test strip drum have been
used. Then replace the batteries and the
test strip drum.
If you use NiMH rechargeable batter-
ies instead of normal batteries, fewer
than 50 more tests can be performed. If
the rechargeable batteries are old and
exhausted, it may not be possible to per-
form even one more test.
When the batteries are so low that
no more tests can be performed, the
battery symbol appears briefly for 3
seconds in the display. Then the meter
turns off. Insert two batteries (new or
recharged).