Accu-Chek Softclix Bedienungsanleitung Seite 63

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 82
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 62
122
Symbols, display messages and troubleshooting
123
Symbols, display messages and troubleshooting
9
If this happens … it means:
 (cont.)The surrounding temperature is
below 0 °C.
Change location. The temperature
must be between +10 °C and
+40 °C. Allow the meter to warm up
slowly.
Condensation has affected the elec-
tronics.
Let the meter dry slowly.
The meter is defective.
Please contact the Accu-Chek
customer care line (see Chapter
14 “Local customer support and
service”).
The meter is turned on, but
the display is blank.
The display is defective.
Please contact the Accu-Chek customer
care line (see Chapter 14 “Local cus-
tomer support and service”).
If this happens … it means:
The clock has stopped
or
the clock is slow.
The meter was exposed to a tem-
perature below –10 °C and the bat-
teries are beginning to freeze. Turn
the meter off. Let it warm up slowly.
Always keep the meter and bat teries
at temperatures between -10 °C
– +50 °C (with test strip drum:
between +2 °C – +30 °C).
When the meter is turned
on, the clock shows 0:00 or
12.00 a.m.
The meter was exposed to a tem-
perature below -10 °C and the bat-
teries are beginning to freeze. Turn
the meter off. Let it warm up slowly.
Always keep the meter and bat teries
at temperatures between -10 °C
– +50 °C (with test strip drum: be-
tween +2 °C – +30 °C).
The time has not been set.
Set the time as described in Chapter
3.4 “Setting the time”.
Seitenansicht 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 81 82

Kommentare zu diesen Handbüchern

Keine Kommentare