10.1.4 Alert A4: ALARM CLOCK
An alert A4: Alarm clock will occur at the time that you have programmed
into the pump.
Press f twice to turn off and confirm the alert.
Put your insulin pump into RUN, if necessary.
Note If you have set the alarm occurrence to EVERY DAY, the
same alert occurs again every 24 hours. Turn the alarm clock
function to OFF, if you do not want additional alarms at that
same time.
10.1.5 Alert A5: PUMP TIMER
To ensure that insulin delivery is optimized, your pump features a pump
timer.
To inspect the pre-set lifetime of your Accu-Chek Spirit pump before you
start using it, please refer to the “Accu-Chek Spirit Pump Protocol”
included in the starter kit. During use, the remaining lifetime in days
until the expiration date of your insulin pump can be reviewed in the
INFORMATION menu (see “Review the time remaining”).
To ensure you are aware of this timing feature, the alert A5:
PUMP TIMER
will remind you 60 days prior to the end of operation to check your
pump’s programmed remaining operation time.
When the PUMP TIMER has reached zero, the
pump will change into STOP and will no longer
operate.
An E5: END OF OPERATION will appear on the
display.