15.3 Technical standards on electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Spirit is intended for use in the electromagnetic environment
specified below. The customer or the user of the Accu-Chek Spirit should assure
that it is used in such an environment.
Immunity test IEC 60601 test Compliance level Electromagnetic
level environment –
guidance
Electrostatic ± 6 kV contact ± 8 kV contact Avoid any contact
discharge (ESD) ± 8 kV air ± 15 kV air with synthetic
IEC 61000-4-2 (IEC 60601-2-24) materials. A relative
humidity of 10%
will allow ESD to be
more likely.
Power frequency 3 A/m 400 A/m 400 A/m Power frequency
(50/60Hz) magnetic fields
magnetic field (IEC 60601-2-24) should be at levels
characteristic of a
IEC 61000-4-8 typical location in a
typical commercial
or hospital
environment.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Spirit is intended for use in the electromagnetic environment
specified below. The customer or the user of the Accu-Chek Spirit should assure
that it is used in such an environment.
Immunity test IEC 60601 Compliance Electromagnetic environment –
test level level guidance
Conducted RF 3 Vrms Not applicable
IEC 61000-4-6 150 kHz to
80 MHz