Start-up graph over the stabilization period
The start-up graph shows changes in the flow rate over the stabilization
time.
Influence of height on delivery accuracy
The maximum deviation of the delivered amount (overall mean percentage
flow error) for U100 insulin is ≤ ±10%* in case your Accu-Chek Spirit
insulin pump is located ≤ 1 m (3.3 feet) above/below the infusion site.
Bolus
For U100 insulin the maximum deviation of a maximum bolus is ≤ ±5%
and the maximum deviation of a minimum bolus is ≤ ±30%**.
** The measurements were made at a medium basal rate of 1.0 U/h according to
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room temperature.
** The measurements were made according to IEC 60601-2-24:1998 with a Disetronic Classic
infusion set PC 16/110 at room temperature.