136
Checking your meter
Farbe: P 021 – Prüfmittelnummer: 201
After the test strip has dropped out, End is displayed twice letter by let-
ter from left to right. The meter then turns off automatically.
If the result is outside the stated concentration range, perform a second
performance check. If the result of the second check is also outside the
concentration range, please check the following points:
Did you take the test strip from the container that the code
chip in the meter came with?
yes
Did you perform the performance check as described above? yes
Did you wipe the tip of the dropper before you allowed the
test strip to draw up control solution?
yes