Measurement and storage conditions
12
12 Measurement and storage conditions
Pay close attention to the following points. They affect the dependability
of your meter as well as the accuracy of your results.
A meter that has been dropped can produce false results.
12.1 Temperature range
O Without the battery the meter may be kept at between -25 °C and
+70 °C.
O With the battery the meter may be kept at between -10 °C and
+50 °C.
At temperatures above +50 °C the battery could leak and
damage the meter.
At temperatures below -10 °C the battery does not have
enough power to keep the internal clock functioning.
O For blood glucose tests and performance checks the permitted
range is from +10 °C to +40 °C.