Symbols, error messages and troubleshooting
13
This error message ... means:
O You bent or moved the test strip dur-
ing the test.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip.
O The test strip drew up blood or con-
trol solution too soon, i.e. before the
drop symbol was flashing.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip. Always wait for the flashing
drop symbol to appear before allow-
ing the test strip to draw up blood or
control solution.