173
Symbols, error messages and troubleshooting
13
This error message ... means:
O The meter cannot read the code chip.
Remove the code chip and re-insert
it into the meter. If the error message
appears again, you cannot use the
code chip and test strips.
O A code chip from another system has
been inserted. The code chip does not
belong to test strips from the
Accu-Chek Go system.
Remove the code chip and insert the
one belonging to the Accu-Chek Go
test strips you are currently using.
O The code chip is defective or is not
properly inserted in the meter.
Remove the code chip and re-insert
it.